Vendor Session - Application of New ESI Ion Trap Time of Flight Mass Spectrometer in the Identification of Isomeric Drugs

Shaun Bilsborough, Shimadzu Biotech


  • LCMS IT TOF mass spectrometer
    • Mass Errors (external calibration) - Average error = 3.5 ppm
    • External Calibration Stability (24 hrs) - No significant drift over 24 hours
    • Fast Polarity Switching: MS - Positive ion/Negative ion
    • Protein ID - Extensive sequence coverage combined with superior mass accuracy and resolution provides high confidence identification
  • Summary
    • Mass accuracy and mass resolution remains constant in all MS modes
    • The thermally stable design of the flight tube leads to improved mass calibration stability over time (no significant drift over 24 hours)
    • Fast polarity switching allows the identification of analytes not identified by single polarity analysis
    • High mass accuracy and resolution combined with multiple MS fragmentation steps provides increased confidence in the identification of small molecules and proteins