Vendor Session - Application of Nanobore LC-MS to the Analysis of Small Drug Molecules

Gary Valaskovic, New Objective


  • Nanospray System Configuration
    • A standard 6-port injection scheme (1 µL volume) is used with in-line filtration (1 µm porosity) to preserve column integrity
    • ESI high voltage is applied pre-column directly to the mobile phase through a platinum wire micro-electrode
    • Pre-column volume is kept to a minimum (= 1 µL) to minimize the gradient delay
  • Peak Parking Gradient
    • The conventional gradient at 250 nL/min went from 10% B to 90% B over a 3.5 minute period of time
    • The gradient delay between pump and column was approximately 3 minutes
  • Nano-Flow Peak Parking
    • Peak parking gradient
    • 250 nL/min to 150 nL/min
    • 1 µL injection
    • 10 pg on-column